| 
DEVICE DESCRIPTION
The PHOENIX CRx Universal Shunt consists of one silicone elastomer
Cruciform 1m slit valve positioned in a Polysulfone housing,
which is enclosed in a cylindrical silicone pumping chamber
attached to a ventricular catheter with a separate 90 cm,
open end distal catheter. A lot number is stamped on the polysulfone
housing. The silicone valve mechanism inside the polusulfone
valve housing is thimble-shaped with a cross-shaped (cruciform)
slit in its distal end. The mid-portion of the cylindrical
silicone body serves as a pumping chamber. There is an integral
polysulfone connector at the Inlet of the valve housing to
facilitate connection of the ventricular catheter. A radiopaque
pressure designation code of one, two or three radiopaque
dots, indicating low, medium, or high pressure, appears on
the connector tube over the pumping chamber.
The pumping Chamber of the shunt assembly is 5.0 mm high
x 21 mm long. The outlet adapter has an outer diameter of
2.3 mm and an inner diameter of 1.3 mm. A ventricular catheter
is bonded to the pumping chamber. There is an attached ventricular
catheter of either 6 cm or 8 cm having an ID of 0.9 mm and
an OD of 2.5 mm. A metal introducer stylet is included. Included
is a separate 90 cm open-end distal catheter with an outer
diameter of 2.5 mm and an inner diameter of 1.2 mm. The Phoenix
CRx Universal Shunt .IS double-packaged, and IS STERILE AND
NON-PYROGENIC unless the Inner pouch IS opened or damaged.
PRESSURE
10 DOTS |
NOMINAL
PRESSURE |
PROXIMAL
CATH
LENGTH |
ORDER NO. |
Very LOW
(no dots) |
20 H2O
mm |
6 cm
8 cm |
CRX620U
CRX820U |
| Low |
40 H2O
mm |
6 cm
8 cm |
CRX640U
CRX840U |
| Medium |
80 H2O
mm |
6 cm
8 cm |
CRX680U
CRX880U |
| High |
120 H2O
mm |
6 cm
8 cm |
CRX690U
CRX890U |
INDICATIONS
This device is intented for use as the principle pressure-regulating
component of a cerebrospinal fluid shunt assembly, used un
the treatment of hydrocephalus.
CONTRAINDICATIONS
This device is not intended for any use other than that indicated.
Ventriculo-atrial shunting is contraindicated in patients
with infection, high risk of infection, pulmonary emboli,
or high atrial pressure associated with congenital heart failure.
Ventriculo-peritoneal shunting is contraindicated in patients
wit peritonitis, infection, high risk of infection, or a history
of the same.
PRECAUTIONS
Aseptic technique is necessary in all phases of the use of
this device.
It is imperative that this device not be handled with bare
fingers or come into contact with lint. Silicone elastomer
is highly electrostatic and therefore susceptible to contamination
by airborne or surface particles, thus minimize exposure of
silicone to air. The presence of these contaminants could
cause adverse tissue reaction. Rubber-shod clamps or washed,
gloved hands are the best means of handling implantable, silicone
devises.
Carefully ligate all connections securely using non-absorbable,
non-metallic 00 sutures. Do not tie ligatures too tightly
or use stainless steel ligatures since silicone elastomer
is easily cut.
Ensure flow of CSF from the distal end of the shunt system
prior to closure.
This device is intended for SINGLE USE ONLY.
Federal Law (USA) restricts this device to sale by or on the
order of a physician.
IMPLANTATION
The following procedure is offered as an aid to the surgeon
and is not intended to replace individual clinical judgment.
Soaking CSF shunt components in an antibiotic solution has
been cited as a valuable element of protocols designed to
reduce shunt infections. A right parieto-occipital approach
is described here, although placement of the scalp incision
may vary with the surgeon's preference and other factors.
1. After induction of general anesthesia, the head is turned
to the left so that the right posterior parietal scalp is
accessible. The scalp, neck and, in cases where the peritoneal
catheter is used, the chest and abdomen are shaved, washed
and then cleaned with an iodine solution. A continuous field
is draped which includes the right posterior parietal portion
of the scalp, the right side of the neck, and when necessary,
the middle third of the anterior chest and upper abdomen.
The exposed skin is covered with an adherent plastic sheet.
A curved incision is planned so that its base is anterior
and slightly inferior. The scalp flap is planned so that a
burr hole, centered on a point approximately 3 cm from the
midline and 7 to 9 cm superior to the external occipital protruberance,
will lie under the middle of the flap. Desirable placement
of the valve is under an area of thick, well nourished scalp,
in a position adjacent to, not directly under, the incision.
2. After the burr hole is made and the ventricular component
inserted.
NOTE:
If a VA shunt is being implanled take care to insure against
air embolism.
|
3. CSF flow
must be verified following each connection, and before
closure.
4. The wound is thoroughly irrigated with antibiotic solution
and during the operation the exposed tissues and shunt components
are covered with lint-free surgical sponges soaked in the
same solution.
5. The valve may be sutured to the pericranium to avoid migrations
of the silicone body with a loop of suture around the inlet
adapter. The wound is closed in two layers; the galea with
a row of interrupted 000 sutures, and the skin with a fine
monofiliament suture.
6. It is the practice of some physicians to use prophylactic
antibiotics starting the day before operation, for the next
three successive days, and to inject the antibiotic into the
ventricle component whenever the ventricular component is
exposed or changed.
THEORY OF OPERATION
The Cruciform progressive resistance,. Differential pressure
valve mechanism, utilized in the Universal Shunt, was designed
by John Holter D.Sc. to control intracranial pressure (ICP)
and reduce overdrainage when shunting cerebrospinal (CSF)
from the ventricles of the brain to the right (Figure 1)

These curves describe the inlet fluid pressure necessary to
drive fluid, influenced by a pulse, through the valve at corresponding
flow rates. They reflect in vitro test results and not the
intracranial pressure of the patient after implantation. The
nominal pressure represents operating pressure at the 2Occ/hr
flow rate. The nominal pressure specifications at other flow
rates are found by following the horizontal lines to the left
side pressure scale. The performance of the valve is such
that flow rate increases gradually with increasing differential
fluid pressure and, conversely, decreases gradually with decreasing
differential fluid pressure. (Figure 1)
WARNINGS
Tumor dissemination. inadequate drainage and over drainage
of cerebrospinal fluid, infection, disconnection, skin erosion,
and bacterial colonization of the valve have been reported
as complications of shunting in general and may be assumed
to be potential complications of the use of Phoenix valves
and shunt systems.
Cardiopulmonary complications such as pulmonary embolism,
cardiactamponade and atrial perforation, have been reported
In the use of ventriculo-atrial (VA) shunts.
STERILIZATION
The product has been sterilized with 100% ethylene-oxide gas.
If the inner package seal is broken or the package has been
opened or is damaged, sterility may be compromised. In this
case do not use the product.
RE-STERILIZA TION
THIS PRODUCT IS RECOMMENDED FOR SINGLE USE ONLY.
PRE-IMPLANT VALVE PREPARATION:
NOTE: For
optimal results, it is recommended by PBC that the procedures,
outlined below, be followed prior to implantation. Failure
to do so may result in substandard valve performance
1. Fill a sterile bowl with sterile physiologic solution.
2. Using aseptic technique, remove the shunt from the package.
3. Remove the inserts from the distal slits.
4. Fill the catheter with sterile saline using a syringe.
5. Ensure that all air has been removed from the lumen, then
clamp the tubing.
6. Immerse all components Into the bowl of sterile physiologic
solution.
7. Check that all of the Air bubbles have been removed from
the tubing and the valve body prior to Implant.
CAUTION:
The presence of air bubbles in. the system will adversely
affect valve function.
Note: Due to the physical characteristics of silicone elastomer,
no shortl-term test of silicone elastomer valves can provide
a precise indication of a valve's pressure designation or
performance According to the FDA, patency and reflux are the
only Issues that can be adequately addressed in a short-term
OR test. For this reason, no OR pressure / flow test is recommended.
For more information, refer to the pamphlet, "Pressure-Flow
Characteristics as They Relate to the treatment of Hydrocephalus,"
available on request from PBC.
PRODUCT WARRANTY
Phoenix Biomedical Corp Warrants that, during manufacture,
packaging and sterilization of Phoenix products, care was
taken to make the devices suitable for use as indicated. Seller
warrants that at the time of sale the goods shall be free
from defects in workmanship and material. The foregoing warranty
is in lieu of all other warranties, expressed or implied,
and no other warranties exist including, without limitation,
any warranty of merchantability or fitness for any other purpose
other than that designated on the product insert sheet. Whether
or not the device is appropriate for use in a particular patient
depends on the nature of the patient's signs and symptoms,
and such determination is the sole responsibility of the physician.
If the goods do not conform to the warranty set forth herein,
seller's sole liability and obligation shall be to replace
the goods or refund the purchase price. In no event shall
seller be liable for property damage, personal injury, or
any consequential damages.
Product Description Subject to changes
|